IR spectroscopy is an ideal method for quantifying benzene at the concentration levels required, and there are several standard methods for this measurement, all of which utilize the distinctive C–H out-of-plane deformation band at around 673 cm-1. While this band is characteristic of benzene, toluene and xylenes have some weak absorption at this frequency that can interfere with the results if high concentrations are present.